FPW90-SL Ultra-Low Refrigerated / Heating Circulator
low temperatures below -50 °C with bath opening for internal and external applications
The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The in-struments provide the entire variety of functions of the professional HighTech Series of circulators.
Order number including voltage version | 9352791.07 |
---|---|
Working temperature range | -90 ... +100 °C |
Temperature stability | ±0.05 °C |
Heating capacity | 3 kW |
Pump capacity flow pressure | 22 ... 26 l/min |
Pump capacity pressure | 0.4 ... 0.7 bar |
Usable bath opening (W x L / D) |
28 x 23 / 20 cm |
Filling volume | 22 l |
Total dimensions
(W × L × H) |
59 x 76 x 116 cm |
Product Features
- VFD COMFORT DISPLAY
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- Keypad for setpoints, warning/safety values and menu functions
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Unique early warning system for low liquid level
- Adjustable high temperature cut-out, visible via display
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
- Active Cooling Control
Technical data for FPW90-SL
Total dimensions (W × L × H) |
59 x 76 x 116 cm |
---|---|
Filling volume | 22 l |
Weight | 190 kg |
Pump connections | M16x1 male |
Cooling Water Connection | G¾ in |
Working temperature range | -90 ... +100 °C |
---|---|
Temperature stability | ±0.05 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 3 kW | ||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||
Pump capacity flow pressure | 22 ... 26 l/min | ||||||||||||||||||||||||||||||
Pump capacity pressure | 0.4 ... 0.7 bar | ||||||||||||||||||||||||||||||
Maximum suction | 0.2 ... 0.4 bar | ||||||||||||||||||||||||||||||
Viscosity max. | 70 cSt | ||||||||||||||||||||||||||||||
Power consumption | 19 A |
Bath cover | integrated |
---|---|
Usable bath opening (W x L / D) |
28 x 23 / 20 cm |
Cooling of compressor | 2-stage Water |
---|---|
Cooling water pressure max. | 6 bar |
to provide full cooling capacity according to catalog values | |
Cooling water temperature range | 20 °C |
---|---|
Cooling water difference pressure | 3.5 bar |
Cooling water consumption | 4.2 l/min |
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Interfaces | Profibus optional |
---|---|
External pt100 sensor connection | integrated |
Integrated programmer | 6x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | VFD |
Temperature setting | Keypad |
9352791.16 | 230V/3PPE/60Hz (Without Plug) (R404A) |
---|---|
9352791.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R404A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). |
Accessory for FPW90-SL
Downloads for FPW90-SL
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureFPW90-SL Benefits
For flammable bath fluid
Classification III (FL) according to DIN 12876-1
ATC3. Calibration.
‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.
Process. Under control.
Full regulation of the dynamics control, access to all important control parameters for individual process optimization.
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures