FPW90-SL Ultra-Low Refrigerated / Heating Circulator
low temperatures below -50 °C with bath opening for internal and external applications
The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The in-struments provide the entire variety of functions of the professional HighTech Series of circulators.
Performance data depends on the mains voltage in the country of use. Please select:
Order number including voltage version | 9352791.07 |
---|---|
Working temperature range | -90 ... +100 °C |
Temperature stability | ±0.05 °C |
Heating capacity | 3 kW |
Pump capacity flow pressure | 22 ... 26 l/min |
Pump capacity pressure | 0.4 ... 0.7 bar |
Usable bath opening (W x L / D) |
28 x 23 / 20 cm |
Filling volume | 22 l |
Total dimensions
(W × L × H) |
59 x 76 x 116 cm |
Product Features
- VFD COMFORT DISPLAY
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- Keypad for setpoints, warning/safety values and menu functions
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Unique early warning system for low liquid level
- Adjustable high temperature cut-out, visible via display
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
- Active Cooling Control
Technical data for FPW90-SL
Total dimensions (W × L × H) |
59 x 76 x 116 cm |
---|---|
Filling volume | 22 l |
Weight | 190 kg |
Pump connections | M16x1 male |
Cooling Water Connection | G¾ in |
Working temperature range | -90 ... +100 °C |
---|---|
Temperature stability | ±0.05 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Performance data depends on the mains voltage in the country of use. Please select:
Heating capacity | 3 kW | ||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||
Pump capacity flow pressure | 22 ... 26 l/min | ||||||||||||||||||||||||||||||
Pump capacity pressure | 0.4 ... 0.7 bar | ||||||||||||||||||||||||||||||
Maximum suction | 0.2 ... 0.4 bar | ||||||||||||||||||||||||||||||
Viscosity max. | 70 cSt | ||||||||||||||||||||||||||||||
Power consumption | 19 A |
Bath cover | integrated |
---|---|
Usable bath opening (W x L / D) |
28 x 23 / 20 cm |
Cooling of compressor | 2-stage Water |
---|---|
Cooling water pressure max. | 6 bar |
to provide full cooling capacity according to catalog values | |
Cooling water temperature range | 20 °C |
---|---|
Cooling water difference pressure | 3.5 bar |
Cooling water consumption | 4.2 l/min |
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Interfaces | Profibus optional |
---|---|
External pt100 sensor connection | integrated |
Integrated programmer | 6x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | VFD |
Temperature setting | Keypad |
9352791.16 | 230V/3PPE/60Hz (Without Plug) (R404A) |
---|---|
9352791.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R404A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). |
Accessory for FPW90-SL
Downloads for FPW90-SL
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureFPW90-SL Benefits

Connection of additional equipment
Stakei connections for solenoid valve, HSP booster pump and HST booster heater

Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function

Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages

Control of the external application
External Pt100 sensor connection for precise measurement and control directly in the external application