FL11006 Powerful recirculating cooler
Models with particularly high power reserves
The powerful FL models are suitable for a wide range of cooling tasks in industrial environments, such as removal of large process heat. 2 variants: Air-cooled (FL) and water-cooled (FLW).
![Icon: Heating option for temperature control instrument](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_asafcl78sl6av2qp4roo8eog4r/-S65-FJPG/icon-heizoption.jpg)
Optional heating function
On request, we also offer our FL recirculating chillers with an additional integrated heating function and other special solutions. Our product experts will be happy to advise you individually in order to design a temperature control unit that is exactly right for your needs. Just give us a call!
Performance data depends on the mains voltage in the country of use. Please select:
Order number including voltage version | see refrigerant variants | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -20 ... +40 °C | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temperature stability | ±0.5 °C | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 60 l/min | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pump capacity pressure | 0.5 ... 6 bar | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
Filling volume | 39 ... 47 l | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
Dimensions
(W × L × H) |
78 x 85 x 148 cm |
Product Features
- Ergonomic design and easy operation
- Splash-proof keypad
- Large, bright LED display
- Permissible temperature in return line +80°C
- Easy filling from the top with hinged protective lid
- Low liquid level protection with optical and audible alarm signal
- Integrated stainless steel bath tanks
- Front drain
- No side vents, instruments can be placed right next to other equipment
- RS232 interface for PC connection
- IP class according to IEC 60529: 21
- Alarm output, potential-free change-over contact (max. 30 VA)
- Pressure Indicator
Technical data for FL11006
Dimensions (W × L × H) |
78 x 85 x 148 cm |
---|---|
Filling volume | 39 ... 47 l |
Weight | 248 kg |
Pump connections | G1¼" male |
Barbed fittings inner diameter | 1" |
Working temperature range | -20 ... +40 °C |
---|---|
Temperature stability | ±0.5 °C |
Setting the resolution of the temperature display | 0.1 °C |
Temperature display resolution | 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Return flow temperature max. | 80 °C |
Performance data depends on the mains voltage in the country of use. Please select:
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 60 l/min | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Pump capacity pressure | 0.5 ... 6 bar | ||||||||||||||||||||||||||||||||||||||||||||||||||||||
Power consumption | 17 A |
Bath tank | Stainless steel |
---|
Cooling of compressor | 1-stage Air |
---|
Sound pressure level | 74 dbA |
---|---|
Classification according to DIN12876-1 |
Classification I (NFL) |
IP Code according to IEC 60529 |
IP 21 |
Pump type | Immersion Pump |
Interfaces | Alarm output / RS232 / Stakei |
---|---|
Temperature control | PID1 |
Temperature display | LED |
Temperature setting | Keypad |
9666110.16 | 230V/3PPE/60Hz (Without Plug) (R449A) |
---|---|
9666110.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R452A) |
9666110.S1.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R449A) |
2 each barbed fittings for tubing 1" inner dia (pumpconnections G1 1/4" male). |
Accessory for FL11006
Downloads for FL11006
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureFL11006 Benefits
![Quality control](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_hevpr7dgr57jnafu373htsrb7e/-S65-FJPG/100-25-proofed.jpg)
100% Checked.
100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
![Environmentally friendly design for temperature control units](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_ps5rbagui16lf9opehta6id43r/-S65-FJPG/green-technology.jpg)
Green technology.
Development consistently applied environmentally friendly materials and technologies.
![Logo Made in Germany](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_faesup1da50v9aao0q9jbvr00g/-S65-FJPG/made-in-germany.jpg)
JULABO. Quality.
Highest standards of quality for a long product life.
![Quick guide for JULABO temperature control units](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_aoao523ptt7pjb16gkdbhpsq35/-S65-FJPG/quickstart.jpg)
Quick start.
Individual JULABO consultation and comprehensive manuals at your disposal.