The new generation of laboratory circulators for routine work and standard tasks
Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.
Overview
Technical data
Accessory & Services
Downloads
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow pressure
15 l/min
Pump capacity pressure
0.35 bar
Viscosity max.
50 cSt
Power consumption
14 A
Bath
Bath tank
Stainless steel
Bath cover
integrated
Usable bath opening (W x L / D)
18 x 13 / 15 cm
Cooling
Cooling of compressor
1-stage Air
Other
Classification according to DIN12876-1
Classification III (FL)
IP Code according to IEC 60529
IP 21
Pump function
Pressure Pump
Pump type
Immersion Pump
Electronics
Interfaces
USB
Temperature control
PID1
Absolute temperature calibration
1 Point Calibration
Temperature display
LED
Temperature setting
Keypad
Available voltage versions
9012707.13
230V/60Hz (Nema N6-20 Plug) (R449A)
9012707.03.chn
230V/50Hz (CN Plug) (R449A)
9012707.04
230V/50Hz (UK Plug Type BS1363A) (R449A)
9012707.05
230V/50Hz (CH Plug Type SEV 1011) (R449A)
9012707.02
115V/60Hz (Nema N5-20 Plug) (R449A)
9012707.03
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9012707.12
200V/50-60Hz (Nema N6-20 Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Very bright display makes it easy to read even from a distance.
Timer. Integrated.
CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.
Connection. Easy.
Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.