Case Study: PRESTO A45 Process system
Cooling a Chemglas 20 l reactor from +20 °C to -10 °C with a PRESTO A45 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A45 Process system with a 20 l glass reactor. The PRESTO A45 is connected to the reactor via two 2 m long metal hoses. The PRESTO A45 Process system has been programmed to cool down from +20 °C to -10 °C.

Test Conditions
Product | PRESTO A45 Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 6 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 45 hr:min | ||||||||||||||||
Test volume | 15 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -10 °C | ||||||||||||||||
Reactor | 20l Chemglas (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.19 K |
TN | 622 s |
TV | 59 s |
XPU | 20 K |
Test Results
The PRESTO A45 Process system cools the reactor from +20 °C to -10 °C in 45 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |