Case Study: MAGIO MX-2500F Refrigerated / heating circulator

Cooling a Asahi 20 l reactor from +20 °C to -20 °C with a MAGIO MX-2500F Refrigerated / heating circulator

Objective

This case study tests the cooling capacity of the JULABO MAGIO MX-2500F Refrigerated / heating circulator with a 20 l glass reactor. The MAGIO MX-2500F is connected to the reactor via two 1 m long metal hoses. The MAGIO MX-2500F Refrigerated / heating circulator has been programmed to cool down from +20 °C to -20 °C.

Test Conditions

Product MAGIO MX-2500F Refrigerated / heating circulator
Cooling capacity (Ethanol)
°C 20 0 -20 -30 -40
kW 2.57 1.87 0.92 0.57 0.37
Heating capacity 3 kW
Pump pressure 0.5 bar
Duration 49 hr:min
Test volume 15 l
Setpoint 1 +20 °C
Setpoint 2 -20 °C
Reactor 20l Asahi (glass)
Tubing length 1 m
Fluid in unit Thermal H10
Fluid in reactor Thermal H10

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz

Control Parameters

XP 0.11 K
TN 200 s
TV 10 s
XPU 3.6 K

Test Results

The MAGIO MX-2500F Refrigerated / heating circulator cools the reactor from +20 °C to -20 °C in 49 min.

0 Min.16 Min.33 Min.50 Min.66 Min.-40 °C-30 °C-20 °C-10 °C0 °C10 °C20 °C30 °C
Set point
Temperature in the application
Temperature in temperature control unit
Compare products