Technical data for FP90-SL
Dimensions and volumes
Total dimensions (W × L × H) |
59 x 76 x 116 cm
|
Filling volume
|
22 l
|
Weight
|
201 kg
|
Pump connections
|
M16x1 male
|
Temperature values
Working temperature range
|
-90 ... +150 °C |
Temperature stability
|
±0.05 °C
|
Setting the resolution of the temperature display
|
0.01 °C
|
Temperature display resolution
|
0.01 °C
|
Ambient temperature
|
+5 ... +40 °C
|
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity
|
3 kW |
Cooling capacity
°C |
20 |
0 |
-20 |
-40 |
-60 |
-80 |
kW
1
|
1.8 |
1.7 |
1.6 |
1.35 |
0.75 |
0.15 |
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
Pump capacity flow pressure
|
22 ... 26 l/min |
Pump capacity pressure
|
0.4 ... 0.7 bar |
Maximum suction
|
0.2 ... 0.4 bar
|
Viscosity max.
|
70 cSt
|
Power consumption |
22 A
|
Bath
Bath cover
|
integrated
|
Cooling
Cooling of compressor
|
2-stage Air
|
Other
Classification
according to DIN12876-1
|
Classification III (FL)
|
IP Code
according to IEC 60529
|
IP 21
|
Electronics
Interfaces
|
Profibus optional
|
External pt100 sensor connection
|
integrated
|
Integrated programmer
|
6x60 steps
|
Temperature control
|
ICC
|
Absolute temperature calibration
|
3 Point Calibration
|
Temperature display
|
VFD
|
Temperature setting
|
Keypad
|
Available voltage versions
9352790N150.07 |
400V/3PNPE/50Hz (Plug 32A CEE) (R404A) |
9352790N150.16 |
208V/3PPE/60Hz (Without Plug) (R404A) |
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). |
Downloads for FP90-SL
We would also be happy to send you our printed catalogs and brochures by post.
Request brochure