CORIO CD-310F Refrigerated / heating circulator

The new generation of laboratory circulators for routine work and standard tasks

Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.

With natural refrigerant

The CORIO CD-310F is available with natural refrigerant. Order No. 9 012 713.N1

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-310F
Order number including voltage version see refrigerant variants
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow pressure 15 l/min
Pump capacity pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4 l
Total dimensions
(W × L × H)
23 x 40 x 65 cm

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • USB connection
  • Integrated pump connection M16×1
  • easy switching between internal and external circulation
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • Space-saving cooling coil design yields more usable space in the bath tank

Callback

Product Options

Other options

Technical data for CORIO CD-310F

Dimensions and volumes
Total dimensions
(W × L × H)
23 x 40 x 65 cm
Filling volume 3 ... 4 l
Weight 25.2 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Refrigerant variant N1 (Environmentally-friendly natural refrigerant)
Order number including voltage version 9012713.N1.03
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.31 0.28 0.22 0.13 0.03
Refrigerant R290
Global Warming Potential 3
Filling weight 30 g
Carbon dioxide equivalent 9.0E-5 t
Refrigerant variant S1
Order number including voltage version 9012713.S1.03
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.31 0.28 0.22 0.13 0.03
Refrigerant R449A
Global Warming Potential 1397
Filling weight 63 g
Carbon dioxide equivalent 0.088011 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 15 l/min
Pump capacity pressure 0.35 bar
Viscosity max. 50 cSt
Power consumption 13 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012713.N1.01 100V/50-60Hz (Nema N5-15 Plug) (R290)
9012713.S1.01 100V/50-60Hz (Nema N5-15 Plug) (R449A)
9012713.N1.02 115V/60Hz (Nema N5-15 Plug) (R290)
9012713.S1.02 115V/60Hz (Nema N5-15 Plug) (R449A)
9012713.N1.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9012713.S1.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9012713.N1.03.chn 230V/50Hz (CN Plug) (R290)
9012713.S1.03.chn 230V/50Hz (CN Plug) (R449A)
9012713.N1.04 230V/50Hz (UK Plug Type BS1363A) (R290)
9012713.S1.04 230V/50Hz (UK Plug Type BS1363A) (R449A)
9012713.N1.05 230V/50Hz (CH Plug Type SEV 1011) (R290)
9012713.S1.05 230V/50Hz (CH Plug Type SEV 1011) (R449A)
9012713.N1.13 230V/60Hz (Nema N6-20 Plug) (R290)
9012713.S1.13 230V/60Hz (Nema N6-20 Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Cool-down time

Medium: Ethanol

Heat-up time

Medium: Thermal

Pump capacity

Medium: Water

Accessory for CORIO CD-310F

Downloads for CORIO CD-310F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CD-310F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 15 benefits

JULABO customer survey

We appreciate if you take part and help us to further improve. The survey takes only 3 minutes and is anonymous. Thank you!

Compare products