Case Study: PRESTO W80 Process system

Heating up of an Rettberg 5 l reactor from +20 °C to +50 °C with a PRESTO W80 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO W80 Process system with a 5 l glass reactor. The PRESTO W80 is connected to the reactor via two 1 m metal hoses. The PRESTO W80 Process system has been programmed to heat up from +20 °C to +50 °C.

Test Conditions

Product PRESTO W80 Process system
Cooling capacity (Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 1.2 1.2 1.2 1.2 1.1 1.1 1.1 0.65 0.1
Heating capacity 1.8 kW
Pump pressure 0.4 bar
Duration 35 hr:min
Test volume 5 l
Setpoint 1 +20 °C
Setpoint 2 +50 °C
Reactor 5l Rettberg (glass)
Tubing length 1 m
Fluid in unit Thermal HL80
Fluid in reactor Thermal HL80

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz

Test Results

The PRESTO W80 Process system heats up the reactor from +20 °C to +50 °C in 35 min.

Set point
Temperature in the application
Temperature in temperature control unit