Case Study: PRESTO A40 Process system
Heating up of an Asahi 20 l reactor from -20 °C to +170 °C with a PRESTO A40 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 20 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m metal hoses. The PRESTO A40 Process system has been programmed to heat up from -20 °C to +170 °C.

Test Conditions
Product | PRESTO A40 Process system |
---|---|
Heating capacity | 2.7 kW |
Pump pressure | 0.4 bar |
Duration | 2:00 hr:min |
Test volume | 18 l |
Setpoint 1 | -20 °C |
Setpoint 2 | +170 °C |
Reactor | 20l Asahi (glass) |
Tubing length | 2 m |
Fluid in unit | Thermal HL60 |
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |