FP51-SL Ultra-Low Refrigerated / Heating Circulator
low temperatures below -50 °C for internal and external applications
The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The instruments provide the entire variety of functions of the professional HighTech Series of circulators.
Performance data depends on the mains voltage in the country of use. Please select:
Order number including voltage version | 9352751.06 | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -51 ... +200 °C | |||||||||||||||||||
Temperature stability | ±0.05 °C | |||||||||||||||||||
Heating capacity | 3 kW | |||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||
Pump capacity flow pressure | 22 ... 26 l/min | |||||||||||||||||||
Pump capacity pressure | 5.8 ... 10.2 psi | |||||||||||||||||||
Usable bath opening (W x L / D) |
7.1 x 4.7 / 7.9 in. | |||||||||||||||||||
Filling volume | 7.5 ... 12 l | |||||||||||||||||||
Dimensions
(W × L × H) |
18.1 x 21.7 x 35 in. |
Product Features
- VFD COMFORT DISPLAY
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- Keypad for setpoints, warning/safety values and menu functions
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Unique early warning system for low liquid level
- Adjustable high temperature cut-out, visible via display
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
- Active Cooling Control
Technical data for FP51-SL
Dimensions (W × L × H) |
18.1 x 21.7 x 35 in. |
---|---|
Filling volume | 7.5 ... 12 l |
Weight | 198.4 lbs |
Pump connections | M16x1 male |
Working temperature range | -51 ... +200 °C |
---|---|
Temperature stability | ±0.05 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Performance data depends on the mains voltage in the country of use. Please select:
Heating capacity | 3 kW | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||
Pump capacity flow pressure | 22 ... 26 l/min | |||||||||||||||||||
Pump capacity pressure | 5.8 ... 10.2 psi | |||||||||||||||||||
Maximum suction | 2.9 ... 5.8 psi | |||||||||||||||||||
Viscosity max. | 70 cSt | |||||||||||||||||||
Power consumption | 16 A |
Bath cover | integrated |
---|---|
Usable bath opening (W x L / D) |
7.1 x 4.7 / 7.9 in. |
Cooling of compressor | 1-stage Air |
---|
Sound pressure level | 63 dbA |
---|---|
Classification according to DIN12876-1 |
Classification III (FL) |
IP Code according to IEC 60529 |
IP 21 |
Interfaces | Profibus optional |
---|---|
External pt100 sensor connection | integrated |
Integrated programmer | 6x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | VFD |
Temperature setting | Keypad |
9352751.06 | 230V/3PPE/50Hz (Without Plug) (R507) |
---|---|
9352751.07 | 400V/3PNPE/50Hz (Plug 16A CEE) (R507) |
9352751.S1.07 | 400V/3PNPE/50Hz (Plug 16A CEE) (R449A) |
9352751.16 | 208V/3PPE/60Hz (Without Plug) (R507) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for FP51-SL
Downloads for FP51-SL
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureFP51-SL Benefits
![Safety level S3 unattended operation of temperature control units](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_mo1akcbiep1hrbe7o3sjhj334j/-S65-FJPG/s3.jpg)
For flammable bath fluid
Classification III (FL) according to DIN 12876-1
![Icon Absolute Temperature Calibration](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_hem750ar7l5fjbcp0mdv0n3a65/-S65-FJPG/atc3.jpg)
ATC3. Calibration.
‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.
![Icon Temperature Control Features](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_mpouftvdah76v14amp7ha2us60/-S65-FJPG/tcf.jpg)
Process. Under control.
Full control of the dynamic, access to all important control parameters for individual process optimization.
![Active Cooling Control](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_rf0abv128p0k18694ea0a34334/-S65-FJPG/acc.jpg)
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures