PRESTO W92 Process system
Highest cooling capacity of 27 kW and heating capacity up to 36 kW
The W92 class process systems are extremely powerful and can also simulate extreme environmental conditions. They are used, for example, in the temperature control of vacuum chambers for component testing in the aerospace industry.
The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.
see refrigerant variants | |
Working temperature range | -92 ... +250 °C |
Temperature stability | ±0.05 ... ±0.2 °C |
Heating capacity | 18 kW |
Pump capacity flow pressure | 0 ... 80 l/min |
Pump capacity pressure | 0.1 ... 3 bar |
Total dimensions
(W × L × H) |
95 x 127 x 190 cm |
Product Features
- Cooling capacity up to 31 kW
- Ambient temperature range +5 °C to +40 °C
- Optional analog connections, RS485, Profibus DP, Modbus
- External Pt100 sensor connection
- Analog connections, RS485, Profibus DP (accessory)
- Built-in 5.7’’ industrial color touchscreen
- Alarm output
- Second external Pt100 sensor connection (accessory)
- Intelligent warning and safety functions
- Pump pressure up to 3 bar, max. flow rate 80 l/min
Technical data for PRESTO W92
Total dimensions (W × L × H) |
95 x 127 x 190 cm |
---|---|
Internal usable expansion volume | 40 l |
Minimal process volume | 28 l |
Active heat exchanger volume | 16 l |
Weight | 870 kg |
Pump connections | M38x1.5 male |
Cooling Water Connection | G¾ in |
Working temperature range | -92 ... +250 °C |
---|---|
Temperature stability | ±0.05 ... ±0.2 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 18 kW | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 0 ... 80 l/min | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pump capacity pressure | 0.1 ... 3 bar | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Viscosity max. | 50 cSt | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Power consumption | 43 A |
Cooling of compressor | 2-stage Water |
---|---|
Cooling water pressure max. | 6 bar |
Max. heat dissipated by unit into cooling water | 50 kW |
to provide full cooling capacity according to catalog values | |
Cooling water temperature range | 10 ... 15 °C |
---|---|
Cooling water difference pressure | 1.5 ... 6 bar |
Cooling water consumption 2 | 43 l/min |
2 Cooling water consumption may vary outside recommended cooling water properties. |
Cooling water temperature range | 5 ... 35 °C |
---|---|
Cooling water difference pressure 3 | 0.5 ... 6 bar |
3 At cooling water temperatures of 25 °C and higher, the minimum differential pressure is 1 bar. |
Sound pressure level | 74 dbA |
---|---|
Classification according to DIN12876-1 |
Classification III (FL) |
IP Code according to IEC 60529 |
IP 21 |
Pump type | Centrifugal Pump (Magnetically coupled) |
Interfaces | Alarm output / Ethernet / Modbus TCP/IP / RS232 / SD memory card / USB |
---|---|
External pt100 sensor connection | integrated |
2nd external Pt100 sensor connection | accessory |
Integrated programmer | 8x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | 5.7" TFT Touchscreen |
Temperature setting | Touchscreen |
9421922.07 | 400V/3PNPE/50Hz (Without Plug) (R404A) |
---|---|
9421922.S1.07 | 400V/3PNPE/50Hz (Without Plug) (R449A) |
9421922.17 | 480V/3P(N)PE/60Hz (Without Plug) (R404A) |
Accessory for PRESTO W92
Downloads for PRESTO W92
We would also be happy to send you our printed catalogs and brochures by post.
Request brochurePRESTO W92 Benefits
Green technology.
Development consistently applied environmentally friendly materials and technologies.
Touch display. Perfect operation.
With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.
Convenience for several users
Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Related Products
Products with similar properties