Case Study: PRESTO A80 Process system
Heating up of an Asahi 20 l reactor from 0 °C to +50 °C with a PRESTO A80 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A80 Process system with a 20 l glass reactor. The PRESTO A80 is connected to the reactor via two 2 m metal hoses. The PRESTO A80 Process system has been programmed to heat up from 0 °C to +50 °C.
![PRESTO A80 process system with Asahi glass reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_6p9hvse4cp4v11j7vgaunumc1a/-S421-FJPG/casestudy-presto-a80-asahi-20-liters-reactor.jpg)
Test Conditions
Product | PRESTO A80 Process system | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
Heating capacity | 1.8 kW | ||||||||||||||
Pump pressure | 0.4 bar | ||||||||||||||
Duration | 1:30 hr:min | ||||||||||||||
Test volume | 18 l | ||||||||||||||
Setpoint 1 | 0 °C | ||||||||||||||
Setpoint 2 | +50 °C | ||||||||||||||
Reactor | 20l Asahi (glass) | ||||||||||||||
Tubing length | 2 m | ||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |