PRESTO A70 Process system

2-stage process circulators for low temperatures

The air-cooled PRESTO A70 is JULABO's most compact 2-stage process system and offers high cooling capacities for lowest temperatures down to -75 °C.

The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
A70
Order number including voltage version see refrigerant variants
Working temperature range -75 ... +250 °C
Temperature stability ±0.01 ... ±0.05 °C
Heating capacity 1.5 kW
Pump capacity flow pressure 0 ... 40 l/min
Pump capacity pressure 1.5 ... 24.7 psi
Total dimensions
(W × L × H)
22.4 x 25.2 x 34.6 in.
Arrange callback

Product Features

  • Analog connections, RS485, Profibus DP (accessory)
  • Space optimized design creates more room directly next to the units
  • External Pt100 sensor connection
  • Built-in 5.7’’ industrial color touchscreen
  • Second external Pt100 sensor connection (accessory)
  • Alarm output
  • Temperature stability ±0.01 °C ... ±0.05 °C
  • Pump pressure up to 1.7 bar, max. flow rate 40 l/min
  • Removable ventilation grid
  • Climate-friendly thanks to natural refrigerant

Technical data for PRESTO A70

Dimensions and volumes
Total dimensions
(W × L × H)
22.4 x 25.2 x 34.6 in.
Internal usable expansion volume 4.8 l
Minimal process volume 2.3 l
Active heat exchanger volume 1.7 l
Weight 308.6 lbs
Pump connections M24x1.5 male
Temperature values
Working temperature range -75 ... +250 °C
Temperature stability ±0.01 ... ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1.5 kW
Cooling capacity
Refrigerant variant N1 (Environmentally-friendly natural refrigerant)
Order number including voltage version 9420701.N1.14
Cooling capacity 1 Medium: Ethanol
°C 20 0 -20 -40 -50 -60 -70
kW 1 1.12 1.06 0.98 0.9 0.69 0.45 0.23
Cooling capacity 2 Medium: Ethanol
°C 20 0 -20 -40 -50 -60 -70
kW 1 0.97 0.91 0.84 0.75 0.55 0.38 0.12
Cooling capacity 3 Medium: Thermal HL90
°C 20 0 -20 -30 -40 -50 -60 -70
kW 1 0.96 0.87 0.8 0.76 0.72 0.61 0.35 0.12
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R449A R170
Global Warming Potential 1397 6
Filling weight 650 g 99 g
Carbon dioxide equivalent 0.90805 t 0.000594 t
Refrigerant variant N2 (Environmentally-friendly natural refrigerant)
Order number including voltage version 9420701.N2.14
Cooling capacity 1 Medium: Ethanol
°C 20 0 -20 -40 -50 -60 -70
kW 1 1.12 1.06 0.98 0.9 0.69 0.45 0.23
Cooling capacity 2 Medium: Ethanol
°C 20 0 -20 -40 -50 -60 -70
kW 1 0.97 0.91 0.84 0.75 0.55 0.38 0.12
Cooling capacity 3 Medium: Thermal HL90
°C 20 0 -20 -30 -40 -50 -60 -70
kW 1 0.96 0.87 0.8 0.76 0.72 0.61 0.35 0.12
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R290 R170
Global Warming Potential 3 6
Filling weight 280 g 99 g
Carbon dioxide equivalent 0.00084 t 0.000594 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Cooling capacity 1 = capacity at minimum pump level, cooling capacity 2 = capacity at maximum pump level

Pump capacity flow pressure 0 ... 40 l/min
Pump capacity pressure 1.5 ... 24.7 psi
Viscosity max. 50 cSt
Power consumption 16 A
Cooling
Cooling of compressor 2-stage Air
Other
Sound pressure level 63 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 20
Pump type Centrifugal Pump (Magnetically coupled)
Electronics
Interfaces Alarm output / Ethernet / Modbus TCP/IP / Profibus optional / REG/EPROG optional / RS232 / RS485 optional / SD memory card / Standby-Input optional / USB
External pt100 sensor connection integrated
2nd external Pt100 sensor connection accessory
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display 5.7" TFT Touchscreen
Temperature setting Touchscreen
Available voltage versions
9420701.N1.14 208V/60Hz (Nema N6-20 Plug) (R449A)
9420701.N2.14 208V/60Hz (Nema N6-20 Plug) (R290)
9420701.N1.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9420701.N2.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9420701.N1.04 230V/50Hz (UK Plug Type BS1363A) (R449A)
9420701.N2.04 230V/50Hz (UK Plug Type BS1363A) (R290)
Contact expert

Heat-up time

Medium: Thermal HL90

Cool-down time

Medium: Thermal HL90

Accessory for PRESTO A70

Downloads for PRESTO A70

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

PRESTO A70 Benefits

Touch display. Perfect operation.

With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.

Convenience for several users

Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable

100 % Cooling capacity

‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures

Intelligent temperature control.

Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.

Show all 19 benefits
Compare products