CORIO C-B19 Open heating bath circulator with stainless steel bath tank
Laboratory circulators for internal temperature application at working temperatures from +20 to +100°C
The open heating bath circulators of the new CORIO series feature high-quality bath tanks made of stainless steel.
Order number including voltage version | 9011419.22 |
---|---|
Working temperature range 1 | +20 ... +100 °C |
Temperature stability | ±0.03 °C |
Heating capacity | 1 kW |
Usable bath opening (W x L / D) |
11.8 x 13.8 / 5.9 in. |
Filling volume | 14 ... 19 l |
Dimensions
(W × L × H) |
15 x 22.8 x 16.5 in. |
1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler. |
Product Features
- Bright, white, easy to read display
- Very quiet
- High-quality bath tanks made of stainless steel
- for internal applications
- Easy operation
- Integrated drain screw (except B5)
- Integrated drain screw
Technical data for CORIO C-B19
Dimensions (W × L × H) |
15 x 22.8 x 16.5 in. |
---|---|
Filling volume | 14 ... 19 l |
Weight | 23.1 lbs |
Working temperature range 1 | +20 ... +100 °C |
---|---|
Temperature stability | ±0.03 °C |
Setting the resolution of the temperature display | 0.1 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler. |
Heating capacity | 1 kW |
---|---|
Viscosity max. | 1 cSt |
Circulation capacity flow rate | 6 l/min |
Circulation capacity pressure | 1.5 psi |
Power consumption | 10 A |
Bath tank | Stainless steel |
---|---|
Usable bath opening (W x L / D) |
11.8 x 13.8 / 5.9 in. |
Cooling coil | accessory |
---|
Classification according to DIN12876-1 |
Classification I (NFL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Pump type | Immersion Pump |
Temperature control | PID1 |
---|---|
Absolute temperature calibration | 1 Point Calibration |
Temperature display | LED |
Temperature setting | Keypad |
9011419.22 | 100-115V/50-60Hz (Nema N5-15 Plug) |
---|---|
9011419.33.chn | 230V/50-60Hz (CN Plug) |
9011419.33 | 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) |
9011419.04 | 230V/50-60Hz (UK Plug Type BS1363A) |
9011419.05 | 230V/50-60Hz (CH Plug Type SEV 1011) |
Accessory for CORIO C-B19
Downloads for CORIO C-B19
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureCORIO C-B19 Benefits
Brilliant.
Very bright display makes it easy to read even from a distance.
Switch on. And off you go.
Intelligent operating concept. Ready for operation with just a few quick and easy steps.
Timer. Integrated.
CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.
Related Products