Case Study: PRESTO A30 Process system
Heating up of an QVF 6 l reactor from +20 °C to +200 °C with a PRESTO A30 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A30 Process system with a 6 l glass reactor. The PRESTO A30 is connected to the reactor via two 2 in. metal hoses. The PRESTO A30 Process system has been programmed to heat up from +20 °C to +200 °C.

Test Conditions
Product | PRESTO A30 Process system |
---|---|
Heating capacity | 2.7 kW |
Pump pressure | 7.3 psi |
Duration | 1:40 hr:min |
Test volume | 5 l |
Setpoint 1 | +20 °C |
Setpoint 2 | +200 °C |
Reactor | 6l QVF (glass) |
Tubing length | 2 in. |
Fluid in unit | Thermal HL60 |
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |