Case Study: PRESTO A40 Process system

Heating up of an Normag 10 l reactor from 0 °C to +100 °C with a PRESTO A40 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 10 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. metal hoses. The PRESTO A40 Process system has been programmed to heat up from 0 °C to +100 °C.

Test Conditions

Product PRESTO A40 Process system
Heating capacity 2.7 kW
Pump pressure 5.8 psi
Duration 1:15 hr:min
Test volume 10 l
Setpoint 1 0 °C
Setpoint 2 +100 °C
Reactor 10l Normag (glass)
Tubing length 2 in.
Fluid in unit Thermal HL40
Fluid in reactor Thermal HL40

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz

Test Results

The PRESTO A40 Process system heats up the reactor from 0 °C to +100 °C in 1:15 h.

Set point
Temperature in the application
Temperature in temperature control unit