Case Study: PRESTO W91tt Process system
Heating up of an Büchiglas 100 l reactor from -50 °C to +20 °C with a PRESTO W91tt Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W91tt Process system with a 100 l glass reactor. The PRESTO W91tt is connected to the reactor via two 2 in. metal hoses. The PRESTO W91tt Process system has been programmed to heat up from -50 °C to +20 °C.
![Case study Presto W92 100-liter Buechiglas reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_rv0d7di8nh6cr7kt54hr5l6s4t/-S421-FJPG/casestudy-presto-w92-buechiglas-100-liters-reactor.jpg)
Test Conditions
Product | PRESTO W91tt Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 36 kW | ||||||||||||||||
Pump pressure | 6.5 psi | ||||||||||||||||
Duration | 2:55 hr:min | ||||||||||||||||
Test volume | 100 l | ||||||||||||||||
Setpoint 1 | -50 °C | ||||||||||||||||
Setpoint 2 | +20 °C | ||||||||||||||||
Reactor | 100l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 in. | ||||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |