Case Study: PRESTO A40 Process system
Heating up of an Asahi 20 l reactor from +100 °C to +170 °C with a PRESTO A40 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 20 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. metal hoses. The PRESTO A40 Process system has been programmed to heat up from +100 °C to +170 °C.
![PRESTO A40 process system with Asahi glass reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_8k3k4pq80d3b39r5uu4cokm15s/-S421-FJPG/casestudy-presto-a40-asahi-20-liters-reactor.jpg)
Test Conditions
Product | PRESTO A40 Process system | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
Heating capacity | 2.7 kW | ||||||||||||||
Pump pressure | 5.8 psi | ||||||||||||||
Duration | 1:15 hr:min | ||||||||||||||
Test volume | 18 l | ||||||||||||||
Setpoint 1 | +100 °C | ||||||||||||||
Setpoint 2 | +170 °C | ||||||||||||||
Reactor | 20l Asahi (glass) | ||||||||||||||
Tubing length | 2 in. | ||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |