Case Study: PRESTO A85t Process system
Cooling a Büchiglas 50 l reactor from +20 °C to lowest possible temperature with a PRESTO A85t Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO A85t Process system with a 50 l glass reactor. The PRESTO A85t is connected to the reactor via two 2 in. metal hoses. The PRESTO A85t Process system cools from +20 °C to -73 °C maximum low temperature.
Test Conditions
Product | PRESTO A85t Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 15 kW | ||||||||||||||||
Pump pressure | 7.3 psi | ||||||||||||||||
Duration | 6:30 hr:min | ||||||||||||||||
Test volume | 50 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -85 °C | ||||||||||||||||
Reactor | 50l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 in. | ||||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.2 K |
TN | 720 s |
TV | 100 s |
XPU | 24 K |