Case Study: PRESTO W55 Process system

Cooling a Büchiglas 50 l reactor from +150 °C to +20 °C with a PRESTO W55 Process system

Objective

This case study tests the cooling capacity of the JULABO PRESTO W55 Process system with a 50 l glass reactor. The PRESTO W55 is connected to the reactor via two 2 in. long metal hoses. The PRESTO W55 Process system has been programmed to cool down from +150 °C to +20 °C.

Test Conditions

Product PRESTO W55 Process system
Cooling capacity (Ethanol)
°C 20 0 -10 -20 -30 -40 -50
kW 15 10 6.5 4 2.5 1.2 0.3
Heating capacity 15 kW
Pump pressure 7.3 psi
Duration 53 hr:min
Test volume 35 l
Setpoint 1 +150 °C
Setpoint 2 +20 °C
Reactor 50l Büchiglas (glass)
Tubing length 2 in.
Fluid in unit Thermal HL60
Fluid in reactor Thermal H20S

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 400V/3PNPE/50Hz

Control Parameters

XP 0.1 K
TN 385 s
TV 21 s
XPU 12 K

Test Results

The PRESTO W55 Process system cools the reactor from +150 °C to +20 °C in 53 min.

0 Min.16 Min.33 Min.50 Min.66 Min.0 °C50 °C100 °C150 °C
Set point
Temperature in the application
Temperature in temperature control unit

JULABO customer survey

We appreciate if you take part and help us to further improve. The survey takes only 3 minutes and is anonymous. Thank you!

Compare products