Case Study: PRESTO A40 Process system
Cooling a Asahi 20 l reactor from +170 °C to +100 °C with a PRESTO A40 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A40 Process system with a 20 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. long metal hoses. The PRESTO A40 Process system has been programmed to cool down from +170 °C to +100 °C.
Test Conditions
Product | PRESTO A40 Process system |
---|---|
Heating capacity | 2.7 kW |
Pump pressure | 5.8 psi |
Duration | 1:05 hr:min |
Test volume | 18 l |
Setpoint 1 | +170 °C |
Setpoint 2 | +100 °C |
Reactor | 20l Asahi (glass) |
Tubing length | 2 in. |
Fluid in unit | Thermal HL60 |
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |