CORIO CP-310F Refrigerated / heating circulator

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

With natural refrigerant

The CORIO CP-310F is available with natural refrigerant. Order No. 9 013 713.N1

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-310F
Order number including voltage version see refrigerant variants
Working temperature range -30 ... +200 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
Refrigerant variant N1 (Environmentally-friendly natural refrigerant)
Order number including voltage version 9013713.N1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.3 0.27 0.21 0.12 0.02
Refrigerant R290
Global Warming Potential 3
Filling weight 30 g
Carbon dioxide equivalent 9.0E-5 t
Refrigerant variant S1
Order number including voltage version 9013713.S1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.3 0.27 0.21 0.12 0.02
Refrigerant R449A
Global Warming Potential 1397
Filling weight 63 g
Carbon dioxide equivalent 0.088011 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4 l
Dimensions
(W × L × H)
23 x 40 x 65 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • easy switching between internal and external circulation
  • Integrated pump connection M16×1
  • Space-saving cooling coil design yields more usable space in the bath tank
  • early warning system for low liquid level

Technical data for CORIO CP-310F

Dimensions and volumes
Dimensions
(W × L × H)
23 x 40 x 65 cm
Filling volume 3 ... 4 l
Weight 25.6 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -30 ... +200 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Refrigerant variant N1 (Environmentally-friendly natural refrigerant)
Order number including voltage version 9013713.N1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.3 0.27 0.21 0.12 0.02
Refrigerant R290
Global Warming Potential 3
Filling weight 30 g
Carbon dioxide equivalent 9.0E-5 t
Refrigerant variant S1
Order number including voltage version 9013713.S1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.3 0.27 0.21 0.12 0.02
Refrigerant R449A
Global Warming Potential 1397
Filling weight 63 g
Carbon dioxide equivalent 0.088011 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Viscosity max. 50 cSt
Power consumption 13 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013713.N1.01 100V/50-60Hz (Nema N5-15 Plug) (R290)
9013713.S1.01 100V/50-60Hz (Nema N5-15 Plug) (R449A)
9013713.N1.02 115V/60Hz (Nema N5-15 Plug) (R290)
9013713.S1.02 115V/60Hz (Nema N5-15 Plug) (R449A)
9013713.N1.33 200-230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9013713.S1.33 200-230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9013713.N1.33.chn 200-230V/50-60Hz (CN Plug) (R290)
9013713.S1.33.chn 200-230V/50-60Hz (CN Plug) (R449A)
9013713.N1.04 200-230V/50-60Hz (UK Plug Type BS1363A) (R290)
9013713.S1.04 200-230V/50-60Hz (UK Plug Type BS1363A) (R449A)
9013713.N1.05 200-230V/50-60Hz (CH Plug Type SEV 1011) (R449A)
9013713.S1.05 200-230V/50-60Hz (CH Plug Type SEV 1011) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CORIO CP-310F

Downloads for CORIO CP-310F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CP-310F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Switch on. And off you go.

Intelligent operating concept. Ready for operation with just a few quick and easy steps.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 16 benefits