CORIO CP-201F Refrigerated / heating circulator
with an enhanced pump performance and a temperature range up to +200 °C
Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.
Order number including voltage version | 9013702.04 |
---|---|
Working temperature range | -20 ... +200 °C |
Temperature stability | ±0.03 °C |
Heating capacity | 2 kW |
Pump capacity flow pressure | 8 ... 27 l/min |
Pump capacity pressure | 0.1 ... 0.7 bar |
Usable bath opening (W x L / D) |
13 x 15 / 15 cm |
Filling volume | 3 ... 4 l |
Total dimensions
(W × L × H) |
44 x 41 x 44 cm |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- powerful and infinitely adjustable pressure pump
- USB connection
- RS232 interface for online communication
- Bath lid and drain tap included
- removable ventilation grid
- Refrigeration unit without side vents
- Class III (FL) according to DIN 12876-1
- easy switching between internal and external circulation
- Integrated pump connection M16×1
- Space-saving cooling coil design yields more usable space in the bath tank
- early warning system for low liquid level
Technical data for CORIO CP-201F
Total dimensions (W × L × H) |
44 x 41 x 44 cm |
---|---|
Filling volume | 3 ... 4 l |
Weight | 24.7 kg |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -20 ... +200 °C |
---|---|
Temperature stability | ±0.03 °C |
Setting the resolution of the temperature display | 0.1 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 2 kW | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||
Pump capacity flow pressure | 8 ... 27 l/min | |||||||||||||||||||||
Pump capacity pressure | 0.1 ... 0.7 bar | |||||||||||||||||||||
Viscosity max. | 50 cSt | |||||||||||||||||||||
Power consumption | 12 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
13 x 15 / 15 cm |
Cooling of compressor | 1-stage Air |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Pump function | Pressure Pump |
Pump type | Immersion Pump |
Interfaces | RS232 / USB |
---|---|
Temperature control | PID1 |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | LED |
Temperature setting | Keypad |
Electronic Timer h:min |
0 ... 999 |
9013702.01 | 100V/50-60Hz (Nema N5-15 Plug) (R134a) |
---|---|
9013702.02 | 115V/60Hz (Nema N5-15 Plug) (R134a) |
9013702.04 | 230V/50-60Hz (UK Plug Type BS1363A) (R134a) |
9013702.05 | 230V/50-60Hz (CH Plug Type SEV 1011) (R134a) |
9013702.33 | 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R134a) |
9013702.33.chn | 230V/50-60Hz (CN Plug) (R134a) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CP-201F
Downloads for CORIO CP-201F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureCORIO CP-201F Benefits
Brilliant.
Very bright display makes it easy to read even from a distance.
Switch on. And off you go.
Intelligent operating concept. Ready for operation with just a few quick and easy steps.
Timer. Integrated.
CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.
Related Products
Products with similar properties