Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
CORIO CP-1001F
|
|
Working temperature range
|
-38 ... +100 °C
|
Temperature stability
|
±0.03 °C
|
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in.
|
Filling volume
|
48 ... 56 l
|
Dimensions
(W × L × H) |
17.7 x 25.2 x 37.4 in.
|
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- powerful and infinitely adjustable pressure pump
- USB connection
- RS232 interface for online communication
- Bath lid and drain tap included
- Removable ventilation grid
- Refrigeration unit without side vents
- Class III (FL) according to DIN 12876-1
- easy switching between internal and external circulation
- Integrated pump connection M16×1
- Space-saving cooling coil design provides more usable space in the bath tank
- early warning system for low liquid level
Technical data for CORIO CP-1001F
Dimensions and volumes
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in.
|
Filling volume
|
48 ... 56 l
|
Weight
|
162.5 lbs
|
Pump connections
|
M16x1 male
|
Barbed fittings inner diameter
|
8/12 mm
|
Temperature values
Working temperature range
|
-38 ... +100 °C |
Temperature stability
|
±0.03 °C
|
Setting the resolution of the temperature display
|
0.1 °C
|
Temperature display resolution
|
0.01 ... 0.1 °C
|
Ambient temperature
|
+5 ... +40 °C
|
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Bath
Bath tank
|
Stainless steel
|
Bath cover
|
integrated
|
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in.
|
Cooling
Cooling of compressor
|
1-stage Air
|
Other
Classification
according to DIN12876-1
|
Classification III (FL)
|
IP Code
according to IEC 60529
|
IP 21
|
Pump function
|
Pressure Pump
|
Pump type
|
Immersion Pump |
Electronics
Interfaces
|
RS232 / USB
|
Temperature control
|
PID1
|
Absolute temperature calibration
|
3 Point Calibration
|
Temperature display
|
LED
|
Temperature setting
|
Keypad
|
Electronic Timer
h:min
|
0 ... 999
|
Available voltage versions
9013708.04 |
230V/50-60Hz (UK Plug Type BS1363A) (R449A) |
9013708.05 |
230V/50-60Hz (CH Plug Type SEV 1011) (R449A) |
9013708.33 |
230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A) |
9013708.33.chn |
230V/50-60Hz (CN Plug) (R449A) |
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CP-1001F
Downloads for CORIO CP-1001F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochure