Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
CORIO CD-B5
Order number including voltage version
9012405.02
Working temperature range
1
+20 ... +150 °C
Temperature stability
±0.03 °C
Heating capacity
1 kW
Pump capacity flow pressure
16 l/min
Pump capacity pressure
4.8 psi
Usable bath opening(W x L / D)
5.9 x 5.9 / 5.9 in.
Filling volume
3.5 ... 5 l
Dimensions
(W × L × H)
9.1 x 15 x 16.1 in.
1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.
Bright, white, easy to read display
Very quiet
easy switching between internal and external circulation
USB connection
High-quality bath tanks made of stainless steel
For internal and external applications
Integrated drain screw (except B5)
Integrated pump connection M16×1
Technical data for CORIO CD-B5
Dimensions and volumes
Dimensions(W × L × H)
9.1 x 15 x 16.1 in.
Filling volume
3.5 ... 5 l
Weight
18.1 lbs
Pump connections
M16x1 male
Barbed fittings inner diameter
8/12 mm
Temperature values
Working temperature range
1
+20 ... +150 °C
Temperature stability
±0.03 °C
Setting the resolution of the temperature display
0.1 °C
Temperature display resolution
0.01 ... 0.1 °C
Ambient temperature
+5 ... +40 °C
1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity
1 kW
Pump capacity flow pressure
16 l/min
Pump capacity pressure
4.8 psi
Viscosity max.
50 cSt
Power consumption
10 A
Bath
Bath tank
Stainless steel
Bath cover
accessory
Usable bath opening(W x L / D)
5.9 x 5.9 / 5.9 in.
Cooling
Cooling coil
accessory
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function
Pressure Pump
Pump type
Immersion Pump
Electronics
Interfaces
USB
Temperature control
PID1
Absolute temperature calibration
1 Point Calibration
Temperature display
LED
Temperature setting
Keypad
Available voltage versions
9012405.01
100V/50-60Hz (Nema N5-15 Plug)
9012405.02
115V/60Hz (Nema N5-15 Plug)
9012405.03
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
9012405.03.chn
230V/50Hz (CN Plug)
9012405.04
230V/50Hz (UK Plug Type BS1363A)
9012405.05
230V/50Hz (CH Plug Type SEV 1011)
9012405.13
230V/60Hz (Nema N5-15 Plug)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Add to quotation request
Contact expert
Heat-up time
Medium: Water
Pump capacity
Medium: Water
Accessory for CORIO CD-B5
Downloads for CORIO CD-B5
We would also be happy to send you our printed catalogs and brochures by post.
Request brochure