Case Study: PRESTO W55 Process system

Heating up of an Büchiglas 100 l reactor from +20 °C to +150 °C with a PRESTO W55 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO W55 Process system with a 100 l glass reactor. The PRESTO W55 is connected to the reactor via two 2 m metal hoses. The PRESTO W55 Process system has been programmed to heat up from +20 °C to +150 °C.

Test Conditions

Product PRESTO W55 Process system
Cooling capacity (Ethanol)
°C 20 0 -10 -20 -30 -40 -50
kW 15 10 6.5 4 2.5 1.2 0.3
Heating capacity 15 kW
Pump pressure 0.5 bar
Duration 1:17 hr:min
Test volume 75 l
Setpoint 1 +20 °C
Setpoint 2 +150 °C
Reactor 100l Büchiglas (glass)
Tubing length 2 m
Fluid in unit Thermal HL60
Fluid in reactor Thermal H20S

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 400V/3PNPE/50Hz

Control Parameters

XP 0.2 K
TN 695 s
TV 85 s
XPU 15 K

Test Results

The PRESTO W55 Process system heats up the reactor from +20 °C to +150 °C in 1:17 h.

0 Min.33 Min.66 Min.100 Min.133 Min.166 Min.200 Min.50 °C100 °C150 °C
Set point
Temperature in the application
Temperature in temperature control unit

JULABO customer survey

We appreciate if you take part and help us to further improve. The survey takes only 3 minutes and is anonymous. Thank you!

Compare products