Case Study: PRESTO A40 Process system
Cooling a Chemglas 10 l reactor from +100 °C to 0 °C with a PRESTO A40 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A40 Process system with a 10 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m long metal hoses. The PRESTO A40 Process system has been programmed to cool down from +100 °C to 0 °C.
Test Conditions
Product | PRESTO A40 Process system |
---|---|
Heating capacity | 2.7 kW |
Pump pressure | 0.4 bar |
Duration | 1:30 hr:min |
Test volume | 10 l |
Setpoint 1 | +100 °C |
Setpoint 2 | 0 °C |
Reactor | 10l Chemglas (glass) |
Tubing length | 2 m |
Fluid in unit | Thermal HL40 |
Fluid in reactor | Thermal HL40 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |