Case Study: PRESTO A30 Process system

Cooling a QVF 6 l reactor from +20 °C to lowest possible temperature with a PRESTO A30 Process system

Objective

This case study tests the lowest temperature of the JULABO PRESTO A30 Process system with a 6 l glass reactor. The PRESTO A30 is connected to the reactor via two 2 m metal hoses. The PRESTO A30 Process system cools from +20 °C to -12 °C maximum low temperature.

Test Conditions

Product PRESTO A30 Process system
Heating capacity 2.7 kW
Pump pressure 0.5 bar
Duration 4:00 hr:min
Test volume 5 l
Setpoint 1 +20 °C
Setpoint 2 -40 °C
Reactor 6l QVF (glass)
Tubing length 2 m
Fluid in unit Thermal HL60
Fluid in reactor Thermal HL60

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz

Test Results

The PRESTO A30 Process system cools the reactor from +20 °C to the lowest temperature of -12 °C in 4:00 h.

Set point
Temperature in the application
Temperature in temperature control unit