TE400 Peltier chiller

Energy-efficient temperature control for external applications

With their robust design and precise temperature control, Peltier chillers are ideal for simple temperature control tasks in laboratories and research facilities. They operate without refrigerants and are therefore very environmentally friendly. At the same time, the reliable chillers contribute to lowering operating costs and reducing the ecological footprint thanks to their energy-efficient operation.

The TE400 uses fewer movable components such as compressors for cooling. This makes it comparatively low-maintenance and less susceptible to wear. The Peltier chiller offers a cooling capacity of up to 400 watts and a pump capacity of up to 7 l/min and impresses with its high temperature stability for a wide range of different applications.

Silent Mode

The TE400 Peltier chiller is equipped with an additional operating mode for use with lower cooling power requirements. The noise level is significantly reduced in Silent Mode, contributing to a more pleasant working atmosphere.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
TE400
Order number including voltage version 9635400.99
Working temperature range -5 ... 40 °C
Temperature stability 0.02 °C
Heating capacity 0.45 kW
Pump capacity flow pressure 7 l/min
Pump capacity pressure 1.2 bar
Filling volume 1 l
Total dimensions
(W × L × H)
27.4 x 41.3 x 40 cm
Arrange callback

Product Features

  • High temperature stability
  • Environmentally friendly operation with low energy consumption
  • Economical operation due to high energy efficiency and low energy consumption
  • Touch display for easy operation

Technical data for TE400

Dimensions and volumes
Total dimensions
(W × L × H)
27.4 x 41.3 x 40 cm
Filling volume 1 l
Weight 24 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -5 ... 40 °C
Temperature stability 0.02 °C
Temperature display resolution 0.1 °C
Ambient temperature 15 ... 40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 0.45 kW
Cooling capacity
Cooling capacity 1 Medium: Water
°C 20 10
kW 1 0.4 0.2
Cooling capacity 2 Medium: Water
°C 20 10
kW 1 0.33 0.19
Cooling capacity 3 Medium: Water Glycol 30:70
°C 20 10 0
kW 1 0.4 0.2 0.05

1 Cooling capacity 1&3: Boost mode, cooling capacity 2: Silent mode. Cooling performance specifications apply for ambient temperature of 20°C. The cooling capacity may decrease if the ambient temperature rises.

Pump capacity flow pressure 7 l/min
Pump capacity pressure 1.2 bar
Power consumption 4.35 A
Other
Sound pressure level 2 70 dbA

2 Silent Mode: 62dB

Electronics
Interfaces RS232
Available voltage versions
9635400.99 100-230V/50-60Hz (Multiple Plugs (EU / US / GB / CH - Plug))

Accessory for TE400

Downloads for TE400

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

JULABO customer survey

We appreciate if you take part and help us to further improve. The survey takes only 3 minutes and is anonymous. Thank you!

Compare products