FC1200T Chiller
with highest pump capacity as well as Pt100 and analog connection
FC recirculating coolers with a particularly powerful pump and Pt100 connection are preferably used for precise temperature control of large-volume external applications. They can be integrated into process control systems via the permanently integrated analog connection.
Order number including voltage version | 9600126.04 | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -10 ... +80 °C | |||||||||||||||||||||
Temperature stability | ±0.2 °C | |||||||||||||||||||||
Heating capacity | 1.2 kW | |||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||
Pump capacity flow pressure | 28 l/min | |||||||||||||||||||||
Pump capacity pressure | 3.5 bar | |||||||||||||||||||||
Filling volume | 8 ... 11 l | |||||||||||||||||||||
Dimensions
(W × L × H) |
46 x 61 x 49 cm |
Product Features
- Adjustable ratio for feed/return temperatures
- Easy filling system located at the front
- Activation of pump for filling
- Bright MULTI-DISPLAY (LED)
- Upper and lower temperature warning functions with interval tone
- Integrated freezing protection and dry-running protection
- Rapid and easy operation via seamless, splash-proof keypad
- Liquid level indication on the front
- Self-test, reciprocal sensor monitoring, pump motor and compressor overload protection
- Complete shut-down with audible signal in case of an alarm
- removable venting grid for simplified removal of dust
- RS232 interface for PC connection
- High temperature stability
- Integrated heater with 1.2 kW capacity
- Connections for external Pt100 sensor
- Analog Connections for programming (0..10V / 0...24mA) and temperature recorder (0V = 0°C)
- Pressure Indicator
Technical data for FC1200T
Dimensions (W × L × H) |
46 x 61 x 49 cm |
---|---|
Filling volume | 8 ... 11 l |
Weight | 67 kg |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -10 ... +80 °C |
---|---|
Temperature stability | ±0.2 °C |
Temperature display resolution | 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 1.2 kW | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||
Pump capacity flow pressure | 28 l/min | |||||||||||||||||||||
Pump capacity pressure | 3.5 bar | |||||||||||||||||||||
Power consumption | 8 A |
Bath tank | Stainless steel |
---|
Cooling of compressor | 1-stage Air |
---|
Sound pressure level | 58 dbA |
---|---|
Classification according to DIN12876-1 |
Classification III (FL) |
IP Code according to IEC 60529 |
IP 21 |
Pump type | Immersion Pump |
Interfaces | Alarm output / Profibus optional / REG/EPROG / RS232 / Stakei / Standby-Input |
---|---|
External pt100 sensor connection | integrated |
Temperature control | PID1 |
Temperature display | 2x LED |
Temperature setting | Keypad |
9600126.03 | 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R134a) |
---|---|
9600126.04 | 230V/50Hz (UK Plug Type BS1363A) (R134a) |
9600126.05 | 230V/50Hz (CH Plug Type SEV 1011) (R134a) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for FC1200T
Downloads for FC1200T
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureFC1200T Benefits
100% Checked.
100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
Green technology.
Development consistently applied environmentally friendly materials and technologies.
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Control of the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Related Products
Products with similar properties