Case Study: PRESTO A40 Process system

Heating up of an QVF 6 l reactor from +20 °C to +200 °C with a PRESTO A40 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 6 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m metal hoses. The PRESTO A40 Process system has been programmed to heat up from +20 °C to +200 °C.

Test Conditions

Product PRESTO A40 Process system
Heating capacity 2.7 kW
Pump pressure 0.5 bar
Duration 1:45 hr:min
Test volume 5 l
Setpoint 1 +20 °C
Setpoint 2 +200 °C
Reactor 6l QVF (glass)
Tubing length 2 m
Fluid in unit Thermal HL60
Fluid in reactor Thermal HL60

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz

Test Results

The PRESTO A40 Process system heats up the reactor from +20 °C to +200 °C in 1:45 h.

Set point
Temperature in the application
Temperature in temperature control unit
Compare products