Case Study: PRESTO A80t Process system
Cooling a Asahi 20 l reactor from +150 °C to +20 °C with a PRESTO A80t Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A80t Process system with a 20 l glass reactor. The PRESTO A80t is connected to the reactor via two 1 m long metal hoses. The PRESTO A80t Process system has been programmed to cool down from +150 °C to +20 °C.

Test Conditions
Product | PRESTO A80t Process system | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||||
Heating capacity | 2.8 kW | ||||||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||||||
Duration | 2:20 hr:min | ||||||||||||||||||||
Test volume | 19 l | ||||||||||||||||||||
Setpoint 1 | +150 °C | ||||||||||||||||||||
Setpoint 2 | +20 °C | ||||||||||||||||||||
Reactor | 20l Asahi (glass) | ||||||||||||||||||||
Tubing length | 1 m | ||||||||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 208V/60Hz |