Technical data for FCW600
Dimensions and volumes
Total dimensions (W × L × H) |
36 x 54 x 49 cm
|
Filling volume
|
6 ... 8 l
|
Weight
|
45 kg
|
Pump connections
|
M16x1 male
|
Cooling Water Connection
|
G¾ in
|
Barbed fittings inner diameter
|
8/12 mm
|
Temperature values
Working temperature range
|
-20 ... +80 °C |
Temperature stability
|
±0.2 °C
|
Temperature display resolution
|
0.1 °C
|
Ambient temperature
|
+5 ... +40 °C
|
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity
|
1.2 kW |
Cooling capacity
°C |
20 |
10 |
5 |
0 |
-10 |
kW
1
|
0.6 |
0.47 |
0.4 |
0.34 |
0.21 |
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
Pump capacity flow pressure
|
20 l/min |
Pump capacity pressure
|
0.5 bar |
Power consumption |
7 A
|
Bath
Bath tank
|
Stainless steel
|
Cooling
Cooling of compressor
|
1-stage Water
|
Cooling water pressure max.
|
6 bar
|
Recommended cooling water properties
to provide full cooling capacity according to catalog values |
Cooling water temperature range
|
20 °C
|
Cooling water difference pressure
|
3.5 bar
|
Other
Sound pressure level
|
51 dbA
|
Classification
according to DIN12876-1
|
Classification III (FL)
|
IP Code
according to IEC 60529
|
IP 21
|
Pump type
|
Immersion Pump |
Electronics
Interfaces
|
Alarm output / Profibus optional / RS232 / Stakei / Standby-Input
|
External pt100 sensor connection
|
not available
|
Temperature control
|
PID1
|
Temperature display
|
2x LED
|
Temperature setting
|
Keypad
|
Available voltage versions
9601060.13 |
230V/60Hz (Nema N6-20 Plug) (R134a) |
9601060.03 |
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R134a) |
9601060.05 |
230V/50Hz (CH Plug Type SEV 1011) (R134a) |
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Downloads for FCW600
We would also be happy to send you our printed catalogs and brochures by post.
Request brochure