PRESTO A85 Process system
Powerhouses among the process system
High cooling capacities enable extremely low temperatures down to -85 °C possible. The high heating capacity, particularly with the A85t and the W85t, provides even more flexibility in the application.
The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.
Order number including voltage version | 9420852.07 | ||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -85 ... +250 °C | ||||||||||||||||||||||||||||||||
Temperature stability | ±0.05 ... ±0.1 °C | ||||||||||||||||||||||||||||||||
Heating capacity | 6 kW | ||||||||||||||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 0 ... 80 l/min | ||||||||||||||||||||||||||||||||
Pump capacity pressure | 0.1 ... 3 bar | ||||||||||||||||||||||||||||||||
Dimensions
(W × L × H) |
61 x 108 x 125 cm |
Product Features
- Cooling capacity up to 2.8 kW
- External Pt100 sensor connection
- Connections for alarm-output, Pt100 external sensor and standby-input
- Analog connections, RS485, Profibus DP (accessory)
- Built-in 5.7’’ industrial color touchscreen
- Second external Pt100 sensor connection (accessory)
- Alarm output
- Heating capacity up to 15 kW
- Temperature stability ±0.05 °C ... ±0.1 °C
- Pump pressure up to 3 bar, max. flow rate 80 l/min
- removable ventilation grid
Technical data for PRESTO A85
Dimensions (W × L × H) |
61 x 108 x 125 cm |
---|---|
Internal usable expansion volume | 7 l |
Minimal process volume | 9.5 l |
Active heat exchanger volume | 5 l |
Weight | 365 kg |
Pump connections | M30x1.5 male |
Working temperature range | -85 ... +250 °C |
---|---|
Temperature stability | ±0.05 ... ±0.1 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 6 kW | ||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 0 ... 80 l/min | ||||||||||||||||||||||||||||||||
Pump capacity pressure | 0.1 ... 3 bar | ||||||||||||||||||||||||||||||||
Viscosity max. | 50 cSt | ||||||||||||||||||||||||||||||||
Power consumption | 18 A |
Cooling of compressor | 2-stage Air |
---|
Sound pressure level | 69 dbA |
---|---|
Classification according to DIN12876-1 |
Classification III (FL) |
IP Code according to IEC 60529 |
IP 20 |
Pump type | Centrifugal Pump (Magnetically coupled) |
Interfaces | Alarm output / Ethernet / Modbus / Profibus optional / REG/EPROG optional / RS232 / RS485 optional / SD memory card / Standby-Input optional / USB |
---|---|
External pt100 sensor connection | integrated |
2nd external Pt100 sensor connection | accessory |
Integrated programmer | 8x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | 5.7" TFT Touchscreen |
Temperature setting | Touchscreen |
9420852.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R507) |
---|---|
9420852.16 | 208-230V/3PPE/60Hz (Without Plug) (R507) |
9420852.06 | 230V/3PPE/50Hz (Plug 32A CEE) (R507) |
Accessory for PRESTO A85
Downloads for PRESTO A85
We would also be happy to send you our printed catalogs and brochures by post.
Request brochurePRESTO A85 Benefits
Touch display. Perfect operation.
With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.
Convenience for several users
Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Intelligent temperature control.
Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.
Related Products
Products with similar properties